Page 152 - master
P. 152

Directions for use for Thoracoscopic 3-Way-Tap



          1.0 General information           3.0.1 Inspect chrome               4.0 Care and Cleaning of
          •  Please read these directions carefully   plating                  individual components
          before using the instrument.      •  Inspect instrument carefully for damage to
          •  New instruments must be washed, rinsed,   chrome plating.         4.1 Components
          dried, inspected and sterilized before using,
          as described below.               Remove any instrument found to be dam-
                                            aged.
          2.0 Sterilization
                                            3.0.2 Function tests                                      Pos. 3: closing sheet,
                                                                                                      art.no.12-06210-00
          2.1 Steam sterilization                                              Not available
                                            Check:                                                   Pos. 4: rotation adapter,
          •  Before sterilizing, individual components                                               art.no.04-11000-04
          are to be cleaned following the instructions   •  Is the instrument functional?
          under point 4.3 below.            •  Can the valve be opened and closed?
          •  All components can be steam sterilized   •  Is the valve complete?                  Pos. 5: O-ring,
                                                                                                 art.no.15-12051-50
          under following conditions:       •  Is the piston oiled and sealed?
          pressure: max. 2 bar              Warning! If any irregularity is found during
                                            the inspection, the defect must be correct-  4.2 Dismantling the
          time: 25 minutes max              ed beforeusing the instrument
                                                                               instrument
          gravity steam (wrapped) / 132º C/ 270º F /   3.1 Directions for use
          15 minutes                                                           •  Hold the valve body, lift the closing plate
                                                                               and slide it to the side.
          Prevac (wrapped) / 132º C / 270º F / 4   3.1.1 Attaching to suction/
          minutes                           irrigation unit

          Gravity Steam (unwrapped/ Flash) / 132ºC/
          270º F / 10 minutes
          Notice: Cold Soak Sterilization is not rec-
          ommended                                                             •  Then push the piston out of the body of
                                                                               the valve using your thumb.
                                              Suction tube
          Testing results indicate that cold soak ster-
          ilization is not  adequate for this product.
          Manufacturer’s  recommended cycle times
          DO NOT always provide a 105 sterility assur-
          ance level.
          Warning! Follow the manufacturer’s in-
          structions for the sterilizer
          2.2 Gas sterilization (ETO)               Irrigation tube
          •  Before sterilizing, individual components
          are to be cleaned following the instructions   •  Connect the silicone tube to the
          under point 4.0 below.            longitudinally directed tube connector as in   4.3 Cleaning instrument
          •  All components can be sterilized with gas.   illustration.        parts
          Warning! Follow the manufacturer’s in-  •  Connect the silicone tube to the
          structions for the sterilizer     transverse directed tube connector as in   MICTEC 3-way-Valves must be taken apart for
                                            illustration.                      cleaning, as described under point 4.2 above.
          2.3 Disinfecting                  •  Be sure to use the appropriate tube   All tubes must be removed from valve!
                                            diameter to prevent the tube from slipping
          •  Before sterilizing, individual components   off connector         Machine washing:
          are to be cleaned following the instructions                         •  All parts can be cleaned in a machine
          under point 4.3 below.                                               •  Parts must be placed in machine so that
          •  All components can be immersed in                                 coating is not damaged.
          disinfectant solutions that have been                                •  Be sure all hollow spaces are also
          approved by the disinfectant manufacturer                            washed and rinsed.
          for endoscopic instruments.                                          •  Water temperature must not go above
          Warning! Follow the manufacturer’s in-                               45ºC to avoid protein coagulation and
          structions for the sterilizer                                        sticking to the valve.
                                                                               •  Follow instructions given by  manufacturer
          3.0 Inspect before each use                                          for proper amount of detergent.
                                                                               •  Follow manufacturer’s instructions for use
                                                                               of machine.



          150                            Colorado, USA / 1.800.821.7496 / www.SontecInstruments.com
   147   148   149   150   151   152   153   154   155   156   157