Important directions for Thoracoscopic Instruments



                                                brush.
                                                •  If you are using an ultrasonic machine use
                                                a good enzymatic cleaner as suggested by
                                                the manufacturer to release the blood and
                                                other organic proteins.
                                                •  After cleaning and rinsing, all parts must
                                                be thoroughly dried.
                                                Sterilization
                                                •  Reassemble parts before sterilizing -
                                                Trocar must remain complete with obturator
                                                inserted.
                                                •  All trocar parts can be sterilized in a
                                                steam autoclave only.
                                                •  Gamma, hot air and plasma sterilization
                                                methods cannot be used as these methods
                                                could cause damage to trocar.
                                                •  Maximum pressure .2 bar
                                                •  time: 25 minutes max
                                                Gravity steam (wrapped) / 132º C/ 270º F /
                                                15 minutes

                                                Prevac (wrapped) / 132º C / 270º F / 4
                                                minutes

                                                Gravity Steam  (unwrapped/ Flash) /  132ºC/
                                                270º F / 10 minutes
                                                Notice: Cold Soak Sterilization is not rec-
                                                ommended
                                                Testing results indicate that cold soak ster-
                                                ilization is not  adequate for this product.
                                                Manufacturer’s  recommended cycle times
                                                DO NOT always provide a 105 sterility assur-
                                                ance level.
                                                •  Temperature: 273 deg. Fahrenheit (134d
                                                C)
                                                •  If not reassembled before sterilizing
                                                and you need to assemble under sterile
                                                conditions in the operating room:
                                                •  If the diameter of the flexible sleeve is
                                                smaller than the outer diameter of obturator,
                                                we recommend that you use the correct size
                                                stretching tool (sterilized).
              Thoracoscopic Flex Trocar         •  Take hold of the obturator head with
                                                middle, index finger and thumb.
              Cleaning Instructions             •  Tip must be in upward position.
                                                •  Take the titanium trocar head at the thread
              •  Disassemble trocar into its separate parts
              - titanium head, obturator and flexible tube.  and turn it to attach.
              •  Clean all parts according to   •  Insert Obturator through the titanium
                                                trocar head.
              manufacturers’ instructions provided with
              your ultrasonic machine.          •  Take up the flexible plastic sleeve at the
              •  All parts should be completely immersed   blunt tip end and use the blunt obturator to
              in cleaning fluid.                position the sleeve at the head of the fixation
              •  Clean for three minutes at a frequency of   screw.
                                                •  Turn until all parts are screwed tight
              35 Khz.
              •  Use only distilled water for rinsing   together.
              whether manual or ultrasonic.
              •  To avoid coagulation of any proteins
              that may still adhere to parts, the water
              temperature must not exceed 113d F (45d
              C).
              •  If you are cleaning by hand, use a soft