Ultrasound cleaning:
              The ultrasound cleaning is to prefer to
              mechanical cleaning. This is especially
              true for the sensitive instruments used for
              endoscopies.
              •  Instruments must be completely
              immersed in cleaning solution.
              •  Instruments must be cleaned for at least 3
              minutes at a frequency of 35 KHz.
              •  Instruments can be rinsed either
              mechanically or by hand but rinsing must
              be thorough and if possible with desalinated
              water.
              •  Follow the manufacture’s instructions for
              the ultrasonic unit.
              •  After cleaning instrument parts must be
              clean to visual inspection. After cleaning
              parts must be sufficiently dried.
              4.4 Care of valve
              After cleaning and drying, all moveable parts
              such as the rotation adaptor and joints must
              be treated  with an appropriate preservation
              based on paraffin oil.
              Warning! The rotation adapter and sliding
              surface of the piston must be greased with
              our special care lubricant. (Art. No. 26905-
              00)
              4.5 Assembly

              The valve is to be put back together by reversing
              the order of operations described in point 4.2

              Then test the valve in the same way as de-
              scribed in 3.0.1 and 3.0.2
              5.0 Maintenance

              Should any  serious damage to the coating
              appear on any part of the valve please send
              the valve to us for repair.